A&D GX-M Series Instruction Manual page 136

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

Display
Error code
Alternate
EC,E00 Communications error
EC,E01 Undefined command error
EC,E02 Not ready
EC,E03 Timeout error
EC,E04 Excess characters error
EC,E06 Format error
EC,E07 Parameter setting error
Full memory
The stored calibration history has reached 50 instances. If
Blink
more is stored, the old history will be deleted. Refer to
"13.Data Memory"
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
An undefined command was received.
Confirm the command.
A received command can not be processed.
Example:
The balance received a "Q" command, but not in the weighing
mode.
The balance received a "Q" command while processing a
RE-ZERO command. Adjust the delay time to transmit a
command.
If the timeout parameter is set to "t-Up1", the balance did
not receive the next character of a command within the time
limit of one second. Confirm the communication.
The balance received excessive characters in a command.
Confirm the command.
A command includes incorrect data.
Example:
The data is numerically incorrect.
Confirm the command.
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.
136