10.Function Switch And Initialization

10-1Permit Or Inhibit

The balance stores parameters that must not be changed unintentionally adjustment data for
accurately weighing, data for adapting to the usage environment, data to control the communications
interface, etc. "A function selection switch" is provided to protect those parameters and it can be used
to select "change prohibited" or "changeable (usable)". By setting to "change prohibited", that function
cannot be entered, so inadvertent change.
"Switch for function selection" has the following five.
"Function table", "Calibration using the internal mass", "Calibration using the external weight",
"Automatic self calibration", "Internal mass correction".
1. Press the ON:OFF
2. While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key
to display p5 .
3. Press the PRINT key. Then the balance displays the function switches.
SAMPLE key ꞏꞏꞏꞏ To select a switch to change the parameter. The selected switch blinks.
RE-ZERO key ꞏꞏꞏꞏꞏ To change the parameter of the switch selected.
PRINT key ꞏꞏꞏꞏꞏꞏꞏ To store the new parameter and return to the weighing mode.
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ To cancel the operation (display
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
Example of GX-M series
key to turn off the display.
0
To inhibit changes. (Can not be used.)
1
To permit changes. (Can be used.)
to the weighing mode
The display shown left indicates the factory settings.
Function table
0
To inhibit changes to the function table.
1
To permit changes to the function table.
Calibration using the internal mass (One-touch calibration)
0
To inhibit calibration using the internal mass.
1
To permit calibration using the internal mass.
Calibration using an external weight
0
To inhibit calibration using an external weight.
1
To permit calibration using an external weight.
Automatic self calibration (Calibration due to changes in temperature)
0
To inhibit automatic self calibration.
1
To permit automatic self calibration.
Internal mass correction
0
To inhibit correction
1
To permit correction
Clr ) . Press the
44
CAL key and return