10-3. Description Of The Class "Environment, Display"

Condition ( Cond )
Cond 0
This parameter is for sensitive response to the fluctuation of a weight value. Used
for powder target weighing, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts. After setting, the balance
displays
Notes
In automatic response adjustment, the
automatically.
With "Hold function (
averaging time.
Stability band width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation
per second is less than the parameter, the balance displays the stabilization indicator and
outputs or stores the data. The parameter influences the "Auto print mode"
5t-b 0 This parameter is used for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b 2 This parameter ignores slight fluctuation of a weight value. Used to prevent a
weight value from drifting due to vibration or drafts.
Note
With "Hold function (
stabilization range.
Hold function ( Hold ) (Animal weighing mode)
This function is used to weigh a moving object such as an animal. When the weighing data
is over the weighing range from zero and the display fluctuation is within the stabilization
range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the
weighing pan, the display returns to zero automatically. This function is available only when
the hold function parameter is set to "
and any weighing unit other than the counting mode is selected. The stabilization range and
averaging time are set in "Condition (
Weighing range
GX-8K
GX-8K2
GF-8K
GF-8K2
GX-12K GX-20K
GX-30K GX-32K
GF-12K GF-20K
GF-30K GF-32K
.
SLOW
Hold
Hold
Cond
GX-10K
2g
GF-10K
20g
1
)" set to "ON (
)", this item is used to set the
1
)" set to "ON (
)", this item is used to set the
" (the animal mode indicator
1
)" and "Stability band width (
Averaging time
Cond 0
2 sec. Faster
4 sec.
Cond 1
8 sec.More accurate
Cond 2
37
.
FAST
weighing speed is selected
ANIMAL
)".
5t-b
Stable range
5t-b 0
5t-b 1
5t-b 2
illuminates)
Small
Big