9. FUNCTION SWITCH AND INITIALIZATION

9-1 Permit or Inhibit

The balance stores parameters that must not be changed carelessly (e.g. Calibration data for
accurate weighing, Data for adapting to the operating environment, Control data for the RS-232C
interface). There are five switches for the purpose of protecting these parameters. Each switch can
select either "permit" or "inhibit". "Inhibit" protects parameters against careless operations.
Switches
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The
balance displays
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
CAL key
p5 .
To select the switch to change the parameter.
To change the parameter of the switch selected.
0:To inhibit changes.
To store the new parameter and return to the weighing mode.
To cancel the operation. (
To return to the weighing mode, press CAL key once again.
1:To permit changes
Clr
is displayed.)