9. Function Switch And Initialization
9-1. Permit Or Inhibit
The balance stores parameters that must not be changed unintentionally (Example:
Calibration data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface). There are five switches for the purpose of
protecting parameters. Each switch can select either "permit" or "inhibit". The "inhibit"
protects parameters against unintentional operations.
Step 1 Press the
Step 2 While pressing and holding the
the
SAMPLE
display
p5
Step 3 Press the
Step 4 Set the switches using the following keys.
SAMPLE
RE-ZERO
key
PRINT
key
CAL
Function table
Calibration using the internal mass (One-touch calibration)
Calibration using an external weight
Automatic self calibration (Calibration due to changes in temperature)
Internal mass correction
key to turn off the display.
ON:OFF
key, press the
.
key. Then the balance displays the function switches.
PRINT
key
To select a switch to change the parameter. The selected switch
blinks.
key
To change the parameter of the switch selected.
To inhibit changes. (Can not be used.)
0
To permit changes. (Can be used.)
1
To store the new parameter and return to the weighing mode.
To cancel the operation and return to the weighing mode.
GX-K series. The display shown left indicates the factory settings.
To inhibit changes to the function table.
0
To permit changes to the function table.
1
To inhibit calibration using the internal mass.
0
To permit calibration using the internal mass.
1
To inhibit calibration using an external weight.
0
To permit calibration using an external weight.
1
To inhibit automatic self calibration.
0
To permit automatic self calibration.
1
To inhibit correction
0
To permit correction
1
key and
PRINT
key to
ON:OFF
31