Electrical Check According To National Regulations - AGFA DRYSTAR 5302 5366 Service Manual

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
2.2.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Machine specific safety instructions and hints for repair work
(4) Insert the power cord at the
printer side; leave it
disconnected at the wall side.
(5) Put one probe at the ground
connector of the mains
connector.
(6) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance in this
case has to be ≤ 0.3 Ohm.
Refer to Figure 5.
In case the electrical resistance of one measurement exceeds the limits (0.2
Ohm without mains cable; 0.3 Ohm with mains cable) check the printer for
loose grounding connections!

Electrical Check according to National Regulations.

(1) Perform the electrical check according to National Regulations.
The electrical checks that have to be performed in certain intervals (e.g. each
12 months) or after certain service actions (repair and/or maintenance)
depend on National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent
regulations.