DD+DIS150.05E
HC309020 : powersupplymodule.powerth
Type
Context
Action
Reason
Cure
Part
HC309022 : powersupplymodule.powermotor
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC309024 : powersupplymodule.powerlogic
Type
Context
Action
Reason
Cure
Part
HC309025 : powersupplymodule.power3v3
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
09
Replace
PSU module defect
Replace
MOD09
Defect
09
Replace
Actuators power cables short circuited
Check actuators power cables
PSU module defect
Replace
MOD09
Defect
09
Replace
PSU module defect
Replace
MOD09
Defect
09
Replace
PSU module defect
Replace
MOD09