DD+DIS150.05E
2.5.4.1
Performing Film Calibrations
NOTE:
The film calibration has to be performed two times for each tray.
(1) Select menu "Service Engineer
Tools - Enter Service Command
session - 6 Calibration - 1 Film
Calibration".
(2) Select the 1
calibration film as shown in Figure
12 is printed and the measured
Dmax is shown.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.69 (XR310)
Measured Dmax
(3) Repeat the film calibration a 2
time.
(4) Repeat the film calibration two
times for the 2
2.5.4.2
Performing an internal Densitometer Calibration
(1) Select menu "Service Engineer Tools - Enter Service Command session - 6.
Calibration - 2. Internal densitometer calibration".
(2) Follow the instructions on the browser pages. Refer to Figure 13.
(3) Enter the measured densities in the browser window.
(4) Select "Accept changes".