4
4.1
4.2
4.3
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 10
Agfa Company Confidential
Maintenance points - Power off
The maintenance with powered-off machine comprises following tasks:
Task
Check of general condition
Check of condition of belts and gears
Visual check of the cables
Cleaning of the suction cups
Cleaning of Electronics
Cleaning of Power Supply Air Inlet
Check and Cleaning of Drum
Cleaning of Thermal print head
Electrical check
Check of general condition
(1) Perform a visual check of
Power- and network cables.
Covers
The Output tray
The upper and lower Input tray.
(2) Disconnect the power and network cable
(3) Remove the left and right cover to prepare the printer for the next steps
(4) Remove the films in the film trays including white plastic sheet
Check of condition of belts and gears
(1) Check all belts and gears for visible wear
Visual check of the cables
(1) Check cables inside the printer: The insulation may not show any sign of
wear or damage.
Printed copies of this document are not controlled.