AGFA DRYSTAR 5302 5366 Service Manual page 408

Tabletop, multi-size imager
Table of Contents

Advertisement

Maintenance
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 18
Agfa Company Confidential
(4) After the measured values are filled in, the density reproduction is
calculated automatically.
(5) At the right hand side of the table is shown, whether the Density
Reproduction is ok or not. In case it is not ok you are requested to
carry out a CDM calibration.
X-rite 341
Select densitometer:
4
Square Measured density Target density Density diff. Check
100%
20
20
90%
47
49,4
80%
73
78,8
70%
104
108,2
60%
133
137,6
50%
163
167
40%
189
196,4
30%
218
225,8
20%
246
255,2
10%
274
284,6
0%
306
314
When entering the measured density values the decimal point appears in a different spot,
e.g. measured value = 0.23, value to enter = 23.0
Density
Lower limit
Upper limit
0
0
0
30
-5
5
45
-7
7
60
-8
8
75
-9
9
90
-10
10
105
-11
11
120
-12
12
135
-12
12
150
-13
13
165
-13
13
180
-14
14
195
-14
14
210
-14
14
225
-15
15
240
-15
15
255
-15
15
270
-16
16
285
-16
16
300
-16
16
315
-16
16
(1)
The measured density values have to be entered without decimal point.
(e.g. measured value = .23, value to enter = 23)
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
Dmax = 314
0
0
-2,4
0
-5,8
0
Density Reproduction =
-4,2
0
-4,6
0
No CDM calibration needed for DT2B film
-4
0
-7,4
0
-7,8
0
-9,2
0
-10,6
0
-8
0
Density difference LinCheck image DT2B film in Ds 5302
20
15
X-rite 310
10
X-rite 331
X-rite 341
5
X-rite 301
Mac Beth TR924
0
Gretag D200-2
-5
-10
-15
-20
0
50
100
150
200
Target Density
Figure 18
DD+DIS149.05E
OK
Lower limit
Upper limit
Density difference
250
300
350
536609af.cdr
Edition 1, Revision 0
2005-09-09