Electrical Check; Basic Protective Earth Test - AGFA DRYSTAR 5302 5366 Service Manual

Tabletop, multi-size imager
Table of Contents

Advertisement

Repair and Service
Machine specific safety instructions and hints for repair work
2.2
2.2.1
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.1 / 6
Agfa Company Confidential

Electrical check

In case of power supply replacements or Maintenance check the protective
earth resistance according to the national regulations.
The electrical check is split up in two steps:
Step
Task
1
Perform the basic protective earth test
2
Perform the electrical check according to
National Regulations.

Basic Protective Earth Test

(1) Use a multimeter that is
capable to measure a
resistance < 1 Ohm
(2) Put one probe at the ground
connector of the mains input
(3) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance has to
be ≤ 0.2 Ohm.
Refer to Figure 4.