AGFA DRYSTAR 5302 5366 Service Manual page 114

Tabletop, multi-size imager
Table of Contents

Advertisement

Repair and Service
Machine specific tools, software tools and auxiliary means
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/12
Agfa Company Confidential
(2) Field 2 consists of 256 small squares, with grey values from 0 to 255.
Each square has inside another smaller square, with a grey level which is
2 grey steps darker than the surrounding square.
Check: all 256 squares must be visible. In addition, a contrast difference
must be visible between the individual square and the small square
inside.
(3) Field 3 contains a flat field with density value 2.0.
Check: no stripes in any direction may be visible.
In case an artifact in the image can be observed, refer to chapter 3.3,
troubleshooting for symptom, reason and cure.