Electrical check
The electrical check is split up in two steps:
Step
Task
1
Perform the basic protective earth test
2
Perform the electrical check according to
National Regulations.
Basic Protective Earth Test
(1) Use a multimeter that is
capable to measure a
resistance < 1 Ohm
(2) Put one probe at the ground
connector of the mains input
(3) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance has to
be ≤ 0.2 Ohm.
Refer to Figure 14.