Known Issues; Software Release 1.8.0_C2; Solved Software Issue - AGFA DRYSTAR 5302 5366 Service Manual

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
1.3

Known Issues

Software 1.8.0_C1 has following known software issues:
In case the Online Diagnosis detects that the Output Module is Causeive and has
to be exchanged, it does not display "MOD 06 - Please replace module" but "E06 -
Please replace module".
The Module pages in the Service Engineer Tools are not yet operational
In case a film jam happens with last film in a package, the printer does not show
"empty film tray" but continues printing with the "protective cover", what leads to
jam with the protective cover.
Automatic film calibration fails in case of a pending maintenance action.
2

Software Release 1.8.0_C2

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an internal
change designed to correct issues, fix defects, or otherwise minor improvements to
performance without changing the software's basic design, structure or functionality.
2.1

Solved Software Issue

Thermal Head calibration could lead to vertical stripes on the film in the range of
Symptom
 0.05 O.D*. units (For comparison: Under best viewing conditions the human eye can
distinguish 0.01 O.D. units).
*O.D. = Optical Density
Software fault
Cause
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
02-2015
Software Releases and Patches
Software Update
11/2005
 SN1156 plus 1143 and SN 1149
On the Agfa HealthCare Library only the
latest software release can be downloaded.
None
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366 /100/300
Repair and Service
Chapter 3.8 / 6
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents