AGFA DRYSTAR 5302 5366 Service Manual page 402

Tabletop, multi-size imager
Table of Contents

Advertisement

Maintenance
4.6
4.7
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 12
Agfa Company Confidential
Cleaning of Power Supply Air Inlet
(1) Clean the power supply air inlet
with a vacuum cleaner.
Check and Cleaning of Drum
(1) Open the top cover
(2) Open the "Hold down bracket"
(3) Open the TH Module
(4) Remove the flat belt
(5) Check the drum for dirt or
damage by turning at the drum
drive pulley.
(6) In case signs of damage are
visible, also check the STI2 test
image printed under step "3.3
Image quality check": A damage
will be visible in the flatfield area
of the STI2 test image in distance
of 111 mm.
(7) Clean the rubber of the drum with
a lint free cloth lightly moistened
with Isopropyl alcohol or Ethanol
(8) Remount the flat belt
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
DD+DIS149.05E
Figure 11
536609df.cdr
Figure 12
Edition 1, Revision 0
2005-09-09