AGFA DRYSTAR 5302 5366 Service Manual page 149

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Machine specific tools, software tools and auxiliary means
Part 1: Recording the DICOM data
(1) Change the IP address of the
printer to any free IP address
in the network (e.g.
192.9.200.202)
(2) Give the service PC the
former IP address of the
printer (e.g. 192.9.200.201)
(3) Start ADVP
(4) Select 'Basic Grayscale' in the
'Start up configuration' screen.
(5) Select 'Application - Comms
Recorder - Start'
(6) In the following window fill in
- Called (Printer) AE Title (this
is the AE title of the printer),
- 'Called (Host) AE Title',
- ADVP Listen port (usually
104: this is the port the
modality sends DICOM
messages),
- Printer IP address (this is the
IP address of the printer)
- printer listen port (usually
104).
(7) Select okay. Now ADVP is
prepared for recording the
communication between
modality and the printer.
Printed copies of this document are not controlled.
(Type 5366/100)
Drystar 5302
Repair and Service
Figure 48: ADVP as Comms
Recorder
Agfa Company Confidential
Chapter 3.2/47

Hide quick links: