DD+DIS150.05E
HC308014 : devilmodule.power.24v
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308015 : devilmodule.power.actuator
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308016 : devilmodule.headvoltage
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Reason4
Cure4
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E308014
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308015
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308016
SeeManual
Connector P1 not connected on TH
Connect P1 on TH
Connector P14 not connected on MOD08 baseboard (Devil or Devius)
Connect P14 on MOD08 baseboard (Devil or Devius)
Connector P3 not connected on MOD08 baseboard (Devil or Devius)
Connect P3 on PSU
PSU module defect
Replace
MOD09