DD+DIS150.05E
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Controls, Connections and Set Up Procedures
Sharpness Adjustment at the Modality
Case 1: Modality sends in page mode:
In case the modality sends the images in page mode (i.e. it does not send
single images to the printer but already the complete, prepared image
information for one film) use Cubic Bell interpolation. It is defined via following
smooth factors:
Possible range
Reasonable values
Default value
Interpolation Smooth Factors
251
245 235
299
less sharp
sharp
Case 2: Modality sends in image mode:
In case the modality sends the images in image mode (i.e. it sends single
images to the printer to let the printer compose the single images for one film)
use Cubic High Res interpolation. It is defined via following smooth factors:
Possible range
Reasonable values
Default value
199
less sharp
Printed copies of this document are not controlled.
Drystar 5302
(Type 5366/100)
Cubic Bell smooth factors
299 (less sharp) to 200 (very sharp)
251 (less sharp) to 235 (very sharp)
245
200
Cubic Bell
too sharp
Figure 13
Cubic HighRes smooth factors
199 (less sharp) to 101 (very sharp)
148 (less sharp) to 138 (very sharp)
143
Interpolation Smooth Factors
148 143 138
sharp
Figure 14
Legend
Limits
Default
251 245 235
299
Reasonable values
536601GJ.CDR
101
Cubic High Res
too sharp
536601GK.CDR
Chapter 1 / 21
Agfa Company Confidential
200