AGFA DRYSTAR 5302 5366 Service Manual page 186

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
Reason5
Cure5
Reason6
Cure6
Reason7
Cure7
Part
Reason8
Cure8
Part
Reason9
Cure9
Part
HC305008 : thmodule.osb.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Part
HC305009 : thmodule.thermalheadheating
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Connector P1 on TH not connected
Connect connector P1 on TH
Connector P14 on MOD08 baseboard (Devil or Devius) not connected
Connect connector P14 on MOD08 baseboard (Devil or Devius)
PSU defect
Replace
MOD09
OSB defect
Replace
OSB
TH defect
Replace
MOD05
Defect
E305008
SeeManual
P1 on OSB not connected
Connect P1 on OSB
P1 on MOD08 baseboard (Devil or Devius) not connected
Connect P1 on MOD08 baseboard (Devil or Devius)
OSB defect
Replace
OSB
Failure
W305009
SeeManual
TH module heater defect
Replace
MOD05
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 30 of 72

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents