AGFA DRYSTAR 5302 5366 Service Manual page 196

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
MODULE : drumdrivemodule
Elabel : No
HC307001 : drumdrivemodule.m1.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Ambient temperature too low when switching on the printer.
Cure: Ensure that ambient temperature is within specifications (+15 ° C to +30 ° C)
HC307002 : drumdrivemodule.s9
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E307001
SeeManual
M1 not connected
Connect M1
Drum drive module defect
Replace
MOD07
Defect
E307002
SeeManual
S9 not connected
Connect S9
Drum drive module defect
Replace
MOD07