DD+DIS150.05E
HC303025 : upperpickupmodule.s18
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
MODULE : lowerpickupmodule
Elabel : Yes
HC304000 : lowerpickupmodule.film.pickupposition
Type
Context
Action
Reason
Cure
HC304001 : lowerpickupmodule.film.bendposition
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W303025
None
Upper pickup passed sensor S18 malfunctioning
check upper pickup passed sensor S18
Upper pickup module defect
Replace
MOD03
Accidental
None
Lower input jam
Remove lower input jam
Accidental
None
Lower input jam
Remove lower input jam
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 19 of 72