DD+DIS150.05E
MODULE : pumpandvalvesmodule
Elabel : No
HC317000 : pumpandvalvesmodule.m9.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Part
HC317001 : pumpandvalvesmodule.v1.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC317002 : pumpandvalvesmodule.v2.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E317000
SeeManual
Motor M9 not connected
Connect motor M9
XS10 not connected
Connect XS10
Pump and valves module defect
Replace
MOD17
Defect
E317001
SeeManual
XS10 not connected
Connect XS10
Pump and valves module defect
Replace
MOD17
Defect
E317002
SeeManual
XS10 not connected
Connect XS10
Pump and valves module defect
Replace
MOD17