AGFA DRYSTAR 5302 5366 Service Manual page 399

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS149.05E
3.3
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Image quality check
(1) In the browser pages select menu
"Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage"
(2) Print following test images on the
largest format available in the film
magazines:
STI1.ZIP
STI2.ZIP
medical_image.ZIP
The test image as displayed
in Figure 4 to Figure 6 will be
printed.
(3) Mark the test films with a
waterproof pen as testfilm before
maintenance
Figure 5
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
Maintenance
Figure 4
Figure 6
Chapter 9 / 9
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents