AGFA DRYSTAR 5302 5366 Service Manual page 250

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
(1) Select in key-operator menu
'6 Calibration - 1 Film Calibration'.
(2) Select upper or lower input tray:
The calibration film as shown in
Figure 2 is printed and the
measured Dmax is shown.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.20 (XR310)
Measured Dmax
(3) Repeat film calibration for the
second input tray.
NOTE:
The calibration is successful, in case the measured Dmax is max. 2% different from
the target Dmax.
The target Dmax depends on the default densitometer which is configured in Key-
operator menu "2 Change Settings - 5 Image Quality - 1 General".
Refer to table below.
Example:
Selected default densitometer:
Measured Dmax DT2B:
Result:
Default Densitometer
X-Rite 310
X-Rite 331
X-Rite 341
X-Rite 301
MacBeth
Gretag D200
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
536701aa.cdr
Default densitometer
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Adjustments and Calibrations
X-Rite 341
3.04 O.D.
Calibration successful
Target Dmax DT2B
3.20 O.D.
3.09 O.D.
3.05 O.D.
3.10 O.D.
3.10 O.D.
3.05 O.D.
Repair and Service
536336as.cdr
Figure 2
Target Dmax DT2C
3.05 O.D.
2.93 O.D.
2.89 O.D.
2.88 O.D.
2.91 O.D.
2.88 O.D.
Chapter 3.6 / 8