DD+DIS150.05E
Additionally following minor software issues are solved:
With software < 2.2.0_C1 the sporadic error message W306015 (CDM intensity
is unstable) appears.
Sporadic "Error System Failure" of certain combinations of RAM and Tyndall
modules .
When booted with a tray open, the printer cannot read an ISO RF tag.
During regular software installation the "Chinese traditional" translation file
does not get unpacked.
After quality control sporadically the (wrong) density values "0.00" are
displayed. When saving the quality control result the correct values are stored.
14

Software Release 2.3.0_C2

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an internal
change designed to correct issues, fix defects, or otherwise minor improvements to
performance without changing the software's basic design, structure or functionality.
NOTE:
As of Software 2.0.0_C2 DRYSTAR 5302 and DRYSTAR AXYS use the same
software.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
02-2015
Software Releases and Patches
Software Upgrade
06/2012
SN ≥ 19776 and additionally in SN 19187,
19304, 19308, 19321, 19328, 19552, 19774
Agfa HealthCare Library
None
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366 /100/300
Repair and Service
Chapter 3.8 / 24