AGFA DRYSTAR 5302 5366 Service Manual page 108

Tabletop, multi-size imager
Table of Contents

Advertisement

Repair and Service
Machine specific tools, software tools and auxiliary means
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/6
Agfa Company Confidential
A copy of the infocounter data is always stored in the battery RAM of the
printer. At software installation the contents of the battery RAM are written to
the hard disk drive.
This guarantees, that the infocounter data are not lost after exchange of the
hard disk drive
To view the infocounter a service PC is required.
Two possibilities exist to get the infocounter on the Service PC:
#
Type of Infocounter analysis:
A
Via Printers Infocounters Report
B
Via Service Command session
We recommend to use for Infocounter Analysis method A, 'Via Printers
Infocounters Report', as this is the fastest way to access the infocounter.
Method B, 'Via Service Command session ' might be useful, when you can
access a PC in the hospital network for a short time, to save the infocounter
file on the printer CF card for analysis on the Service PC.
Printed copies of this document are not controlled.
(Type 5366/100)
Drystar 5302
DD+DIS150.05E
Refer to ...
2.2.1
2.2.2
Edition 1, Revision 0
2005-09-09