AGFA DRYSTAR 5302 5366 Service Manual page 261

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
Overview of the Afssaps / EUREF calibration procedure:
DRYSTAR 5302 calibration procedure for Afssaps/EUREF
Check calibration of densitometer. Details see section 2.5.1.
Check configured default densitometer. Details see section 2.5.2.
YES
Reset CDM, TH, CFC calibrations. Details see section 2.5.3
Perform following calibrations. Details see section 2.5.4:
1. Film calibration for mammo trays (2 x)
2. CDM calibration
3. Film calibration for mammo trays (1 x)
4. Print Head calibration (1x)
5. Film calibration for mammo trays (2 x)
Perform density reproduction check (DRC). Details see section 2.5.5.
Print TG18 QPC pattern. Measure 18 density values.
Enter values in 'EUREF&AFSSAPS GSDF check.xls'. Details see 2.5.6.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Densitometer
calibration ok?
YES
Used densitometer
= default densitometer
in the printer?
YES
Printer installation date
= today?
No
DRC ok?
YES
Deviations < 10%?
YES
Finished
CFC_calibration_5302.cdr
Figure 11
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Adjustments and Calibrations
Perform densitometer
No
calibration acc. to densitometer
manufacturer instructions.
Adjust densitometer in: 'Key-operator menu -
No
2 Change settings – 5 Image Quality
- 1 General settings'
No
No
Perform CFC calibration.
Details see section 2.5.7.
Repeat up to 3 times
if required.
Agfa Company Confidential
Chapter 3.6 / 19