DD+DIS150.05E
HC302003 : lowertraymodule.tagreader
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC302007 : lowertraymodule.s2
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC302016 : lower.filmcal.required
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W302003
None
Tag reader not connected
Connect tag reader
Tag reader defect
Replace
Tag reader
Failure
W302007
None
S2 not connected
Connect S2
Lower tray module defect
Replace
MOD02
Failure
W302016
None
Lower tray needs calibration
Perform lower tray calibration
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 11 of 72

Hide quick links: