AGFA DRYSTAR 5302 5366 Service Manual page 405

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS149.05E
4.9.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
(4) Insert the power cord at the
printer side; leave it
disconnected at the wall side.
(5) Put one probe at the ground
connector of the mains
connector.
(6) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance in this
case has to be ≤ 0.3 Ohm.
Refer to Figure 15.
In case the electrical resistance of one measurement exceeds the limits (0.2
Ohm without mains cable; 0.3 Ohm with mains cable) check the printer for
loose grounding connections!
Electrical Check according to National Regulations.
(1) Perform the electrical check according to National Regulations.
The electrical checks that have to be performed in certain intervals (e.g. each
12 months) or after certain service actions (repair and/or maintenance)
depend on National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent
regulations.
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
Maintenance
536631JN.cdr
Figure 15
Chapter 9 / 15
Agfa Company Confidential