DD+DIS150.05E
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Controls, Connections and Set Up Procedures
Important remark: In changing parameter values this does not automatically
ensure that the value is applied.
Two situations can be distinguished
• In case the modality does not send a parameter, the value set in the
printer is applied to the modality print job.
• In case the modality sends a parameter value, even when an empty
value (" ") is sent, this value still has priority above the value set in
the printer. So, the value sent by the modality has higher priority then
the value set in the printer, and therefore it is always preferred to set
the required parameter value at the side of the modality.
If this cannot be done, an overrule flag can be set for the respective
parameter in the 'DICOM Host profile'. Setting this overrule flag can
only be done by editing the 'DICOM Host profile' via browser.
Refer to section 6.4.4.
Printed copies of this document are not controlled.