Software Release 1.9.1_C1; Solved Software Issues - AGFA DRYSTAR 5302 5366 Service Manual

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
7

Software Release 1.9.1_C1

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an internal
change designed to correct issues, fix defects, or otherwise minor improvements to
performance without changing the software's basic design, structure or functionality.
7.1

Solved Software Issues

Software 1.9.1_C1 solves following software issues:
Sporadic input film jams for films from the upper tray, mainly 14" films:
Symptom 1
Error message at the local display:
Film jam! Open upper input tray and clear film path
Error message in the infocounter: HC314003
The force for driving the film towards the thermal head is sporadically too low.
Cause
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
02-2015
Software Releases and Patches
Software Update
04/2008
≥ SN 5723 plus 5660, 5665, 5667, 5669,
5672, 5676, 5677, 5679, 5683, 5684, 5717,
5719, 5721
On the Agfa HealthCare Library only the
latest software release can be downloaded.
None
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366 /100/300
Repair and Service
Chapter 3.8 / 13
Agfa Company Confidential