DD+DIS238.06E
General safety note, in case of power ON and cover removed:
DANGER:
Risk of electric shock when working at the device with opened cover and
power ON.
Only work at the powered device with opened cover for service purposes, if this is
absolutely unavoidable.
Before working at the powered device with opened cover for service purposes,
first identify the high voltage wires and connectors in the circuit diagram.
if possible, cover high voltage wires and connectors with insulating material.
greatest possible care must be taken, never to get in contact with high voltage wires
or connectors.
General safety note to check tension before working at the powered-off device:
DANGER:
High voltage. Risk of electric shock.
Before working at the device with opened covers, perform following steps to ensure the
device is de-energized:
Switch off the power at the main power switch.
Prevent that the device can be switched on by other persons. Example: Put a sign
"Don't switch on" in local language to the main switch.
Use an appropriate tester and confirm that the device is de-energized (0V!).
Observe additional safety notes at the device and in the specific service documents.
For devices with capacitors (e.g. X-Ray generator): Confirm by measuring the
voltage at the capacitors, that the capacitors are discharged.
For devices with batteries for power supply (e.g. mobile X-ray unit): Observe special
safety notes in the related service manual.
Static discharge at electrical components:
CAUTION:
Static discharge! Electrical components may be destroyed:
For the repair on electrical components, wear a grounding strap
(Order number: CM+9 9999 0830 0) around the wrist and connect the other end of this
strap on a grounded conducting metal piece.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
04-2014
Generic Safety Directions for HealthCare Imaging Products
DOCUMENT CONTROL NOTE:
Generic Safety Directions
Page 20 of 30
Agfa Company Confidential