AGFA DRYSTAR 5302 5366 Service Manual page 38

Tabletop, multi-size imager
Table of Contents

Advertisement

Controls, Connections and Set Up Procedures
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 1 / 4
Agfa Company Confidential
#
Step
Printer settings:
11.
Adjust Film calibration (optional)
12.
Select Default densitometer (optional)
13.
Set Film view (optional)
14.
Adjust Beeps on empty tray (optional)
15.
Adjust behavior for unreadable RF
Tag (optional)
16.
Enter site data
Film calibration
17.
Perform a film calibration for both film
trays
Printer - Modality connection
18.
Let the modality be set-up
19.
Set-up logical printers (optional)
20.
Print diagnostic and/or test images
21.
Adjust image settings (optional)
Acceptance Test
22.
Perform acceptance test
Completion of installation
23.
Instruct the Customer in operation of
the printer
24.
Save settings on backup CF card
Printed copies of this document are not controlled.
Drystar 5302
(Typ 5366/100)
DD+DIS150.05E
Task
refer to .....
Code
APPL
section 4.1
APPL
section 4.2
APPL
section 4.3
APPL
section 4.4
APPL
section 4.5
APPL
section 4.6
INS
section 5
SIC
section 6.1
SIC
section 6.2
SIC
section 6.3
APPL
section 6.4
APPL
section 7
APPL
section 8.1
INS
section 8.2
Edition 1, Revision 0
Okay
2005-09-09

Hide quick links:

Advertisement

Table of Contents