AGFA DRYSTAR 5302 5366 Service Manual page 176

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
HC304002 : lowerpickupmodule.film.homeposition
Type
Context
Action
Reason
Cure
HC304003 : lowerpickupmodule.film.vacuum
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Part
HC304004 : lowerpickupmodule.filmpickup
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Accidental
None
Lower input jam
Remove lower input jam
Accidental
None
Lower input jam
Remove lower input jam
vacuum system failing
clean vacuum tubes , filter , vacuum cups ......
Lower pickup module defect
Replace
vacuum pump
Accidental
None
Lower input jam
Remove lower input jam
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 20 of 72

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents