AGFA DRYSTAR 5302 5366 Service Manual page 266

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
2.5.4.3
Performing a Film Calibration after the internal Densitometer Calibration
NOTE:
The film calibration after the internal densitometer calibration has to be performed one
time for each tray.
(1) Select menu "Service Engineer
Tools - Enter Service Command
session - 6 Calibration - 1 Film
Calibration".
(2) Select the 1
calibration film as shown in Figure
14 is printed and the measured
Dmax is shown.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.69 (XR310)
Measured Dmax
(3) Repeat the film calibration for the
nd
2
2.5.4.4
Performing a Print Head Calibration followed by two Film Calibrations
NOTE:
The print head calibration has to be performed 2 times, if films of width 10" and 14" are
available.
(1) Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl -
Instructions
alcohol. The last cleaning action must be carried out in one way.
(2) In the browser pages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 3. Print Head Calibration'.
(3) Follow the instructions on the browser pages. Refer to Figure 15.
(4) Perform two film calibrations for both trays. Refer to section 2.5.4.1.
(5) Repeat the print head calibration, if films of width 10" and 14" are available.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
st
input tray: The
536736az.cdr
Default densitometer
input tray.
DOCUMENT CONTROL NOTE:
Adjustments and Calibrations
DRYSTAR 5302
Type 5366/100
Repair and Service
Figure 14
Chapter 3.6 / 24
Agfa Company Confidential