Sw Patch Installation - AGFA DRYSTAR 5302 5366 Service Manual

Tabletop, multi-size imager
Table of Contents

Advertisement

Maintenance
5
5.1
5.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 16
Agfa Company Confidential
Maintenance points - Power on
The maintenance with powered-on machine comprises following tasks:
Task
Check of safety switches

SW patch installation

Film calibration
Density Reproduction Check
Test image evaluation
Host system image quality check
Check of safety switches
(1) Switch on the printer. Wait till it
shows the READY screen (takes
approx. 7 minutes)
(2) Open top cover
(3) Check display message: It must
show "Close top cover"
SW patch installation
(1) Install the latest SW patch, in case a recommended SW patch for the
printer is available. Refer to chapter 3.8, Software Releases, Patches.
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
DD+DIS149.05E
Details: see...
5.1
5.2
5.3
5.4
5.5
5.6
536609fa.cdr
Figure 16
Edition 1, Revision 0
2005-09-09

Hide quick links: