DD+DIS150.05E
HC301004 : uppertraymodule.s11
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC301005 : uppertraymodule.s24
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC301006 : uppertraymodule.takeoversensors.working
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W301004
None
S11 not connected
Connect S11
Upper tray module defect
Replace
MOD01
Failure
W301005
None
S24 not connected
Connect S24
Upper tray module defect
Replace
MOD01
Defect
E301006
SeeManual
S11 and S24 not connected
Connect S11 and S24
Upper tray module defect
Replace
MOD01