AGFA DRYSTAR 5302 5366 Service Manual page 189

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
MODULE : outputmodule
Elabel : Yes
HC306000 : outputmodule.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
HC306001 : outputmodule.cdm.init
Type
Context
Action
Reason
Cure
Part
HC306002 : outputmodule.intensity.scan
Type
Context
Action
Reason
Cure
HC306003 : outputmodule.outputsensor.hi
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E306000
SeeManual
XS2 not connected
Connect XS2
XS3 not connected
Connect XS3
Failure
W306001
None
Output module defect
Replace
MOD06
Accidental
None
Internal film jam
Remove film jam
Accidental
None
Output film jam
Remove film jam
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 33 of 72