DD+DIS150.05E
MODULE : eboxfanmodule
Elabel : No
HC320000 : eboxfanmodule.eboxfan.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC320020 : eboxfanmodule.eboxfan.shortcircuit
Type
Context
Action
Reason
Cure2
Part
MODULE : bothtraysmodule
Elabel : No
HC321001 : pickupmodule.alltrays.disabled
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E320000
SeeManual
EBox fan module not connected
Connect EBox fan module
EBox fan module defect
Replace
MOD20
Defect
E320020
SeeManual
EBox fan module defect
Replace
MOD20
Defect
E321001
SeeManual
All inputtrays have detected a mallfunction