DD+DIS150.05E
HC305006 : thmodule.powerlogic
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Reason4
Cure4
Reason5
Cure5
Part
Reason6
Cure6
Part
Reason7
Cure7
Part
HC305007 : thmodule.headvoltageactuator
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Reason4
Cure4
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E305006
SeeManual
Connector P3 on PSU not connected
Connect P3 on PSU
Connectors P2 on TH not connected
Connect connector P2
Connector P1 on TH not connected
Connect connector P1 on TH
Connector P14 on MOD08 baseboard (Devil or Devius) not connected
Connect connector P14 on MOD08 baseboard (Devil or Devius)
PSU defect
Replace
MOD09
OSB defect
Replace
OSB
TH defect
Replace
MOD05
Defect
E305007
SeeManual
Connector P3 on PSU not connected
Connect P3 on PSU
Connectors P2 and P3 on OSB not connected
Connect connectors P2 and P3
Connectors P3 on TH not connected
Connect connector P3 on TH
Connector P2 on TH not connected
Connect connector P2 on TH