Error 30-154: Sorter Nvram Failure - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 30-154: SORTER NVRAM Failure

8F2919
Page
248 of 326
Messages
Sorter: Internal
Communications Failure
(Alarm sounds)
Description
This error occurs when the MICROCONTROLLER detects an error when attempting to read
or write from NVRAM. The error is cleared when power is cycled on the IMAGER.
Result of Error
The PRINTER is able to print, but films might not go to the correct BIN.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
Sorter microcontroller failed
to read/write to NVRAM