Error 21-932, 22-932, Or 23-932: Supply Controller Communications Retry - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07
Error 21-932, 22-932, or 23-932: SUPPLY CONTROLLER Communications
8F2919
Page
Retry
171 of 326
Messages
None
(No alarm)
Description
Note
UPPER DRAWER (U) = 21. MIDDLE DRAWER (M) = 22, LOWER DRAWER (L) = 23.
This error occurs when the MCS tries to communicate with the MICROCONTROLLER but
receives no response. The error is cleared when power is cycled on the IMAGER.
Note
Error 2x-931 is generated when this problem is detected after 5 successive communications
attempts.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
MCS failed to communicate
with xxxxx Supply
Status Messages and Error Codes
DICOM
SEVERITY
NORMAL
3-Notify

Advertisement

Table of Contents
loading

Table of Contents