DIAGNOSTICS
30JUL07
1
To check operation of the SORTER SENSOR, select Diagnostics>MCS>Sorter.
8F2919
Page
2
To do calibration of the SORTER SENSOR:
24 of 326
a
b
c
3
To display the output of the SORTER SENSOR with the results of the last calibration:
a
b
4
After completing the diagnostic, set the SERVICE SWITCH in the user position and
restart the IMAGER.
Set the SERVICE SWITCH in the "Service" mode.
In the top right corner of the screen, click [Run].
The calibration will run and indicate OK. The results will display on the "Monitor
Sensor Charts". You must use the "charts" to determine if a failure occurred. For
example, if the EMITTER does not cause enough light at the DETECTOR, the
"Sensor Reading" will be -1 and the "Sensor LED Current" will be 255.
To end monitoring, on the lower part of the screen, click [Stop].
On the "Monitor Sensor" area of the screen, click [Run].
To stop monitoring, click [Stop].