Error 20-913: DRE Data Transfer Error Notification

8F2919
Page
115 of 326
Messages
None
(No Alarm)
Description
This error occurs when the wrong number of pixels is transferred between the DRE and the
DPB in the allotted time frame. The faulty image will be printed. The job will be failed, placed
in the undelivered queue, and rescheduled for delivery.
Note
Error 20-915 is generated when this problem is detected on 3 successive films.
Result of Error
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.