Error 33-355: Exposure Transport Dsp Communications Failure - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 33-355: EXPOSURE TRANSPORT DSP Communications Failure

8F2919
Page
279 of 326
Messages
Exposure Transport: Internal
Communications Failure
(Alarm sounds)
Description
This error occurs when the MCS tries to communicate with the DSP but receives no response
in 5 attempts. The error is cleared when power is cycled on the IMAGER.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
MCS failed to communicate
with Exposure Transport DSP
after multiple attempts
Status Messages and Error Codes
DICOM
SEVERITY
ELEC DOWN
0-Fatal