Error 29-947: Densitometer Fiducial Reference Diagnostics Warning - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 29-947: DENSITOMETER Fiducial Reference Diagnostics Warning

8F2919
Page
247 of 326
Messages
None
(No alarm)
Description
During a calibration print and during DENSITOMETER diagnostics, a reading from the
FIDUCIAL SENSOR (S17) is taken with the light source turned on. This occurs when the
reading is outside of the optimal range (approaching the range limit).
Note
Error 29-927 is generated when this reading is outside the range limit.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
Densitometer fiducial sensor
reference reading outside of
optimal range with light
source on
Status Messages and Error Codes
DICOM
SEVERITY
NORMAL
3-Notify