Error 29-932: Densitometer Communications Retry - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 29-932: DENSITOMETER Communications Retry

8F2919
Page
245 of 326
Messages
None
(No alarm)
Description
This error occurs when the MCS tries to communicate with the MICROCONTROLLER but
receives no response.
Note
Error 29-931 is generated when this problem is detected after 5 successive communications
attempts.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL