DIAGNOSTICS
30JUL07
Error 21-922, 22-922, or 23-922: RF TAG Diagnostics Failure
8F2919
Page
167 of 326
Messages
xxxxx Supply: Internal
Diagnostics Failure - RF Tag
(Alarm sounds)
Description
Note
UPPER DRAWER (U) = 21. MIDDLE DRAWER (M) = 22, LOWER DRAWER (L) = 23.
When the RF TAG READER is to run diagnostics and returns a diagnostics failure, the MCS
commands a retry. This error occurs if the MICROCONTROLLER indicates a diagnostics
failure after 3 attempts. The error is cleared when power is cycled on the IMAGER.
Result of Error
The system cannot print from this SUPPLY DRAWER. The SPOOLER is able to receive and
queue jobs, and print from any other available SUPPLY DRAWER.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
xxxxx Supply: RF Tag reader
failed self diagnostics
Status Messages and Error Codes
DICOM
SEVERITY
BAD SUPPLY
1-Alert
MGZ