Error 20-918: DRE Data Transfer Too Slow. Fatal Timing Error Detected

8F2919
Page
119 of 326
Messages
Internal Image Data Transfer
Failed
(Alarm sounds)
Description
This error occurs when the rate of image data transfer between the DRE and DPB is
inadequate to feed the EXPOSURE TRANSPORT SUBSYSTEM on 3 successive films. A
faulty image is printed on each failure. Jobs associated with the faulty image will be failed,
placed in the undelivered queue, and rescheduled for delivery. The error is cleared when the
IMAGER power is cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs, but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
2
Install a new DATAPATH USB CABLE from the DRE to the OPTICS MODULE.
3
Install a new HARD DRIVE. See the ADJUSTMENTS AND REPLACEMENTS,
DRIVE - DRE
LOCAL PANEL
AY.
LOG
Image transfer failure after
multiple attempts. Transfer
rate inadequate.