Error 12-080: Software Update Error - Kodak DryView 6800 Diagnostics Manual

Laser imager
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

DIAGNOSTICS
30JUL07

Error 12-080: Software Update Error

8F2919
Page
103 of 326
Messages
None
(No alarm)
Description
The LOCAL PANEL cover screen was notified that the software update was not successfully
completed. The error is logged, and it cannot be cleared until the system is "re-ghosted".
Result of Error
The system is unusable because the failed software update could have left the system in any
state. It is unknown if diagnostic quality images can be generated. Images will not be printed.
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
Software Update Failure
Status Messages and Error Codes
DICOM
SEVERITY
NORMAL
1-Alert

Advertisement